Skip to product information
1 of 1

2-Wire Programmable Transmitter IS (Output 4 to 20mA + HART Protocol)

Regular price
$189.00
Regular price
Sale price
$189.00
  • Input: RTD (platinum, copper, nickel), thermocouple
  • High measurement accuracy
  • 2, 3, 4-wire connection
  • 4 to 20mA output, HART Protocol
  • Range programmable
  • Offset and linearity re-calibration
  • Surge & full pulse protection
  • Input and output electrical isolated (1500V AC)
  • Display unit with powerful programming functions optional
Sku: 61221

Package: 5 pieces

Shipping: 2-3 weeks

2-Wire Programmable Transmitter IS (Output 4 to 20mA + HART Protocol)
2-Wire Programmable Transmitter IS (Output 4 to 20mA + HART Protocol)

$189.00

  • Description
  • Specifications
  • Manual and Resources

This transmitter supports RTD and thermocouple inputs, linearizes the input signals and outputs 4mA to 20mA or signals by HART protocol with accuracy ±0.1%FS of selected range. The DIP switch design makes configuration easily done on site as well as range and error programmable by computer software.

The surge and full pulse protection, input and output electrical makes the transmitter work stable and avoid the damage from the thunder or power supply voltage fluctuations from the high-power consumption equipment startups and stops. The low power consumption design decreases the drift from the temperature by the heat dissipation and improves its long-term stability.

The option of probe selections provides the convenience to the user for the whole measurement system. The calibration of transmitter and probe as a system provides the unbroken traceability chain to the data with uncertainty. The annual calibration service can guarantee the long-time use with reliable data.

Input Max. offset

100%

RTD Input

Pt50, Pt100, Pt200, Pt500, Pt1000

Ni50, Ni100, Ni120, Ni1000

Cu50, Cu100

Cable resistance per wire

50Ω per lead

Sensor current

≤0.3mA

Effect of sensor cable resistance (3-wire)

<0.001Ω/Ω

Linear resistance

0Ω to 20000Ω

Thermocouple input

B, E, J, K, N, R, S, T, C, D

Internal cold end compensation

Pt1000, -50°C to 105°C

External cold end compensation

Pt100, -200°C to 850°C

Cable resistance per wire

10KΩ per lead maximum

Output Current

4 to 20mA/20mA to 4mA optional

Linear current output

3.8mA to 20.75mA programmable by Hart protocol

Load (@ current output)

≤(Vsupply-12)/0.02 Ω

Output load stability

≤0.01% of span/100 Ω

Output sensor error detection

Circuit open: ≤3.6mA

Circuit short: 20.75mA to 21.6mA configurable

Output updating time

100ms

Power supply

12V to 30V DC

Internal power dissipation

43mW to 0.65W

Response time (programmable)

0 to 60s

Voltage drop

12V DC

Warm-up time

3min

Programming

Loop power

Accuracy

0.1%FS selected range

Signal dynamics, input

24 bit

Signal dynamics, output

16 bit

Effect of supply voltage change

<0.002% of span/ VDC

EMC immunity influence

<±0.5% of span

Operating temperature

-40°C to 85°C

Operating relative humidity

<95%rh (non-condenation)

Operating relative humidity

<95%rh (non-condenation)

Operating relative humidity

<95%rh (non-condenation)

Protection grade

IP67/IP00

Calibration temperature

22°C to 28°C

Temperature coefficient

0.008% FS /1℃

Dimensions/ Weight approx.

Φ44mm x 24mm/ 46g

Wire size

≤2.0mm2 stranded wire

Vibration

2 to 100Hz, 4g

ATEX

Ex ia IIC T4 NEPSI

You may also like